Table II: Compounds tested and found not to cross-react with the test as a 1 mg/ml concentration in urine.

The following compounds do not cross-react with Amphetamine:

Acetaminophen
Actylsalicylic Acid
Amikacin
Amitriptyline
Ampicillin
Arterenol
Aspartame
Atropine Sulfate
Benzoic Acid
Benzoylecgonine HCl
Caffeine
Chlorpheniramine
Chlorpropmazine HCl
Cimetidine
Codeine
Deoxyephedrine
Dextromethorphan
Diazepam
Diethylpropion
Diphenylhydantoin
Doxylamine
Ecgonine HCl
Ecgonine Methyl Ester
Glucose
Histamine
Hydrocodone
Hydromorphone
Indomathacin
Ketoprofen
Levorphanol
D - 9 THC
11-nor-D-9-carboxy-THC-9-COOH
Meperidine
Methylphenidate
Methadone
Methaqualone
Morp. Glucuronide
Morphine Sulfate
Oxazepam
Oxycodone
Pendimetrazine
Penicillin G
Pentobarbital
d-Propoxyphene
Hydrochlorthiazide
Propanol
Phencyclidine
Phenobarbital
Phentermine
Phenylpropanolamine
L-Phenylephrine
Quinine
Ranitidine
Sodium Salicylate
Tryptophan
Tetracycline
Tetrahydrozoline
Theophylline
Thioridazine
Triflouperazine

Back to Top

The following compounds do not cross-react with Marijuana:

Acetaminophen
4-Acetamidophenol
Acetylsalicyclic acid
Amikacin
Ampicillin
d,l-Amphetamine
Amitriptyline
Arterenol
Aspartame
Atropine sulfate
Benzoic Acid
Benzoylecgonine
Caffeine
Camphor
Chloroquine
Chlorpheniramine
Chlorpromazine HCl
Cocaine Hydrochloride
Cocaine
Cimetidine
Cortisone
Deoxyepinophrine
Dextromethorphan
Diazepam
Digitoxin
Digoxin
Ecgonine Hydrochloride
Ecgonine Methyl Ester
Ephedrine
Epinephrine
Gentisic acid
Glucose
Guaiacol
Histamine
Hydrochlorothiazide
Hydrocodone Hydromorphone
Homatropine
Imipramine
Isoproterenol
Ketamine
Lidocaine
d-Phencyclidine (PCP)
Meperidine
Methadone
Methaqualone
Methylpehnidate
Morphine glucuronide
Morphine sulfate
Morphine
Naloxone
Neomycin
Niacinamide
Oxazepam
Perphenazine
Penicillin G
Phencyclidine
Phenobarbital
Phenylethylamine-a
Phenylpropanolamine
Promethazine
Pseudophedrine
Rantidine
Salycyclic acid
Secobarbital
Tetracycline
Tetrahydrozoline
Theophylline
Thiridazine
Triflouperazine
Tryptophan

Back to Top

The following compounds do not cross-react with Cocaine:

Acetaminophen
Actylsalicylic Acid
Amikacin
Amitriptyline
Ampicillin
Arterenol
Aspartame
Atropine Sulfate
Benzoic Acid
Benzoylecgonine HCl
Caffeine
Chlorpheniramine
Chlorpropmazine HCl
Cimetidine
Codeine
Deoxyephedrine
Dextromethorphan
Diazepam
Hydromorphone
Indomathacin
Ketoprofen
Levorphanol
D -9-THC
Diethylpropion
Dephenylhydantoin
Doxylamine
Ecgonine HCl
Ecgonine Methyl Ester
Glucose
Histamine
Hydrocodone
11-nor-D -9-carboxy-THC-9-COOH
Meperidine
Methylphenidate
Methadone
Methaqualone
Morp. Glucuronide
Morphine Sulfate
Oxazepam
Oxycodone
Pendimetrazine
Penicillin G
Pentobarbital
d-Propoxyphene Hydro-
chlorothiazide
Propanol
Phencyclidine
Phenobarbital
Phentermine
Phenylpropanolamine
L-Phenylephrine
Quinine
Ranitidine
Sodium Salicylate
Tryptophan
Tetrahydrozoline
Theophylline
Thioridazine
Trifluoperazine

Back to Top

The following compounds do not cross-react with Opiates:

Acetaminophen
Acetylsalicylic Acid
Amikacin
Amitriptyline
Ampicillin
Arterenol
Aspartame
Atropine Sulfate
Benzoic Acid
Benzoylecgonine HCl
Caffeine
Chlorpheniramine
Chlorpropmazine HCl
Cimetidine
Deoxyephedrine
Dextromethorphan
Diazepam
Diethylpropion
Diphenylhydantoin
Doxylamine
Ecgonine HCl
Ecgonine Methyl Ester
Glucose
Histamine
Indomathacin
Ketoprofen
Levorphanol
D - 9 THC
11-nor-D-9-carboxy-THC-9-COOH
Methylphenidate
Methadone
Methaqualone
Morphine Sulfate
Oxazepam
Pendimetrazine
Penicillin G
Pentobarbital
d-Propoxyphene-
Hydrochlorothiazide
Propanol
Phencyclidine
Phenobarbital
Phentermine
Phenylpropanolamine
L-Phenylephrine
Quinine
Sodium Salicylate
Tryptophan
Tetracycline
Tetrahydrozoline
Theophylline
Thioridazine
Trifluoperazine

Back to Top

The following compounds do not cross-react with Phencyclidine (PCP):

Acetylsalicylic Acid
Amikacin
Amitriptyline
Ampicillan
Arterenol
Aspartame
Atropine Sulfate
Benzoic Acid
Benzoylecgonine HCl
Caffeine
Chlorpheniramine
Chlorpropmazine HCl
Cimetidine
Codeine
Deoxyephedrine
Dextromethorphan
Diazepam
Diethylpropion
Dephenylhydantoin
Doxylamine
Ecgonine HCl
Ecgonine Methyl Ester
Glucose
Histamine
Hydrocodone
Hydromorphone
Indomathacin
Ketoprofen
Levorphanol
D 9-TC
11-nor-D-9-carboxy-THC-9-COOH
Meperdine
Methylphenidate
Methadone
Methaqualone
Morp. Glucuronide
Morphine Sulfate
Oxazepam
Oxycodone
Pendimetrazine
Penicillin G
Pentobarbital
d-Propoxyphene-
Hydrochlorthiazide
Propanol
Phencyclidine
Phenobarbital
Phentermine
Phenylpropanolamine
L-Phenylephrine
Quinine
Ranitidine
Sodium Salicylate
Tryptophan
Tetracycline
Tetrahydrozoline
Theophylline
Thioridazine
Triflouperazine

Back to Top

  1. Accuracy: The accuracy of the DrugCheck 5 was first tested in urine samples by D.F.E. and subsequently in a clinical trial of urine samples submitted to a NIDA certified laboratory. In both cases the laboratories used EMIT II as their screening procedure. All positive samples by either screening method were confirmed by GC/MS. The data was combined and the results are summarized as follows:

3.1 Amphetamine (AMP)

  Syva EMIT II Positive Syva EMIT II Negative
DrugCheck 5 Positive 188 1
DrugCheck 5 Negative 0 340

When compared to EMIT II the relative sensitivity between positive samples was 100% . The relative specificity between negative samples was 99.7%. The concordance of the combined data with respect to EMIT II was 99.8%.

Back to Top

3.2 Marijuana (THC)

 

Syva EMIT II Positive

Syva EMIT II Negative

DrugCheck 5 Positive 297 4
DrugCheck 5 Negative 4 259

When compared to EMIT II the relative sensitivity between positive samples was 98.67% . The relative specificity between negative samples was 98.48%. The concordance of the combined data with respect to EMIT II was 98.58%.

Back to Top

3.3 Cocaine (BEG)

 

Syva EMIT II Positive

Syva EMIT II Negative

DrugCheck 5 Positive 194 1
DrugCheck 5 Negative 0 212

When compared to EMIT II the relative sensitivity between positive samples was 100% . The relative specificity between negative samples was 99.55%. The concordance of the combined data with respect to EMIT II was 99.75%.

Back to Top

3.4 Opiates (OPI)

  Syva EMIT II Positive Syva EMIT II Negative
DrugCheck 5 Positive 172 0
DrugCheck 5 Negative 0 190

When compared to EMIT II the relative sensitivity between positive samples was 100% . The relative specificity between negative samples was 100%. The concordance of the combined data with respect to EMIT II was 100%.

Back to Top

3.5 Phencyclidine (PCP) (MET)

  Syva EMIT II Positive Syva EMIT II Negative
DrugCheck 5 Positive 179 14
DrugCheck 5 Negative

0

262

When compared to EMIT II the relative sensitivity between positive samples was 100% . The relative specificity between negative samples was 94.9%. The concordance of the combined data with respect to EMIT II was 96.92%.

Back to Top

Page1 (Table of Contents) - 2-3-4-5-6